HRC 2 BASIC

Harga : Rp 0.00

Brand : IKA

Stok : In Stock





HRC 2 BASIC

HRC 2 Basic is a The compact refrigerated and heating circulator HRC 2 basic is renowned for its high energy efficiency, a working temperature powerful infinitelyrange from -20°C to 100°C and a variable pressure and suction pump. The on-demand and speed-controlled compressor reduces noise and operating costs.

The high-quality Isolation of the 4 l bath allows fast heating and reduces heat input when cooling.
Using the external temperature control circuit is possible for up to 2.5 l thermofluid without a refill.
The integrated hopper and drain valve ensure a safe and clean handling of the thermofluid. The easy-to-open front flap allows an easy cleaning of the cooling unit’s air filter.There is a drain valve on the front side of the bath. A hose can be connected to empty the bath, so the user does not get in direct contact with the thermofluid.

  • Speed-controlled refrigeration system with 400 W cooling capacity
  • Speed-controlled pump make of PEEK: (0.5 bar; 22 l/min)
  • 1500 W heating capacity (1200 W @115V)
  • Digital level indicator
  • Safety classification III (FL) according to DIN 12876
  • Adjustable temperature safety circuit

TECHNICAL DATA

Appliance typeCirculation thermostat
Class designation acc. DIN 12876III
Identification according to DIN 12876FL
Cooling agentR134a
Cooling agent quantity230 g
Cooling agent pressure max.20 bar
Heat output1500 W
Cooling capacity (@20°C)400 W
Cooling capacity (@10°C)370 W
Cooling capacity (@0°C)320 W
Cooling capacity (@-10°C)240 W
Cooling capacity (@-20°C)130 W
Working temperature-20 - 100 °C
Operating temperature min.-20 °C
Temperature displayyes
Temperature stability DIN 12876 (@+70°C)0.1 ±K
Temperature stability DIN 12876 (@-10°C)0.1 ±K
Temperature controlPT 100
Working temperature sensorPT 100
Safety temperature sensorPT1000
Working temperature displayLED
Safety temperature displayLED
Temperature stability DIN 128760.1 ±K
Display resolution0.1 K
Set temperature resolution0.1 ±K
Warning function opticalyes
Warning function acousticyes
Warning function excess temperatureyes
Adjustable safety circuit0 - 110 °C
Sub-level protectionyes
Over-level protectionyes
Filling volume1.4 - 4.0 l
Pump typePressure- / suction pump
Pump capacity adjustableyes
Pump pressure max. (0 liters discharge flow)0.5 bar
Pressure pump (suction side) (0 liter flow)0.25 bar
Flow rate max. (0 bar back pressure)21 l/min
Pump connectionM16x1
Calibration optionyes
Technical data complies with the standardDIN 12876
Permissible ON time100 %
Dimensions (W x H x D)220 x 475 x 525 mm
Weight32.8 kg
Permissible ambient temperature5 - 32 °C
Permissible relative humidity80 %
Protection class according to DIN EN 60529IP 21
RS 232 interfaceyes
USB interfaceyes
Voltage230 / 115 V
Frequency50/60 Hz
Power input1800 W